![]() TM 11-6665-245-12
CHAPTER 2
OPERATING INSTRUCTIONS
2-1. Operator Controls (Figure 2-1)
Refer to Table 2-1 for operator controls and indicators.
Table 2-1.
Operator's Controls and Indicators
CONTROL, INDICATOR, OR
FUNCTION
CONNECTOR
Indicates battery condition or counts per minutes in four ranges depen-
Indicating Meter
ding on selector switch position.
Selector Switch
Turns radiacmeter on and off. Changes internal circuits and indicating
meter to check batteries and measure alpha radiation as follows:
(6 position rotary switch)
Meter Indication
Sw. Pos.
Radiacmeter is off (all other positions: equipment is
OFF
on).
Batteries are good if needles in BATT area of meter
BATT
scale.
Alpha radiation is measured from .0 to 1 million counts
1M
per minute (CPM).
Same as position 1M except that measurement range
100K
is 0 to 100,000 counts per minute (CPM).
Same as position 1M except that measurement range
10K
is 0 to 10,000 counts per minute (CPM).
Same as position 1M except that measurement range
1K
is 0 to 1,000 counts per minute (CPM).
Meter Reset (momentary
Reset indicating meter rapidly to zero.
push button switch)
Applies battery power to small lamp inside indicating meter to illuminate
Light Switch (momentary
indicating meter face.
push button switch)
Headphone Connector
Used to connect CPM electrical signal to earphones.
Change 2
2-1
|