|
|
![]() TM 11-6665-209-20
2-5. TROUBLESHOOTING
The Troubleshooting Table below lists the common malfunctions which may be found
during operation or maintenance of the Radiac Set.
This manual cannot list all malfunctions that may occur, not all tests or inspections
and corrective actions. If a malfunction is not listed or is not corrected by performing
listed corrective actions, notify your Supervisor and evacuate to higher maintenance
for repair.
TABLE 2-3.
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
1. WATER SEEPS INTO BATTERY COMPARTMENT.
Step 1. Remove battery compartment cover by loosening three captive screws
and check the rubber gasket on the mounting panel. REPLACE DEFECTIVE
GASKET.
Remove two screws securing handle to cover and check both gaskets.
Step 2.
REPLACE DEFECTIVE GASKET.
2. RANGE SWITCH IS LOOSE OR BINDING AT ANY SETTING.
Step 1.
Check operation of switch at each position to check for looseness or
binding. If switch binds, loosen the setscrew gradually until the switch
turns easily and stops at each setting. If the switch is loose, tighten
setscrew so the switch stops at each setting.
Do not use excessive force to tighten setscrew. Damage may result.
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |