TM 11-6625-537-14-1
Table 4-4, Organizational Troubleshooting
Probable cause
Corrective action
Malfunction
Defective lamps, or defective fuse F1.
1. Decimal indicator lamps do not
Replace defective lamp(s) (para 4-
1 la ). If no lamps light, replace
light.
fuse F1.
Check for a slight buzzing sound: if
2. CALIBRATE control has no effect
Defective chopper.
on VOLTS meter needle.
there is no buzzing, chopper must
be replaced by general support
maintenance personnel.
instrument to  remove
3. VOLTS meter needle deflects more
Clean
Dirt in circuit board connectors or in
than one-tenth of full scale when
switch contacts causing excessive
controls are set as follows:
electrical leakage.
RANGE
500
NULL
.1
Voltage readout dials
49000
4. VOLTS meter needle deviates from
ZERO ADJ control improperly set.
Adjust ZERO ADJ control R227
zero when controls are set as
.
follows:
RANGE
.5
NULL
.01
Voltage readout dials
o
S e c t i o n IV. ORGANIZATIONAL MAINTENANCE OF ELECTRONIC
V O L T M E T E R ME-202( * )/U
4-10. Replacement of Fuse
a. Connect the electronic voltmeter to a 115-
volt source. If the electronic voltmeter is wired for
I f t h e e l e c t r o n i c v o l t m e t e r i s  completely
230-volt operation, connect to a 230-volt source.
i n o p e r a t i v e , the fuse is probably defective.
b. Mechanically zero VOLTS meter control
Replace the defective fuse with a new one. If the
with mechanical adjustment screw on meter case.
new fuse blows when power is applied, repair at
c. Set ON switch to ON and allow at least a
higher maintenance level is required. Replace the
20 minute warmup.
fuse as follows:
d. Set controls as follows:
.
a. Turn the fuseholder cap on the rear panel
(1) RANGE.5.
counterclockwise to unlock the cap.
(2) NULL .01.
b. Pull out the fuseholder cap and the defective
(3) Voltage readout dials 0.
fuse.
e. Adjust ZERO ADJ, R227, for zero VOLTS
c. Remove the defective fuse and replace it
meter deflection. R227 is in the lower right corner
with a new one.
of the printed circuit board on the left side of the
4.11. VOLTS Meter Zero Adjustment
cabinet when facing the front panel and is ac-
The electrical zero of the VOLTS meter may
c e s s i b l e through the ventilation holes in the
require
adjustment as  indicated by the
cabinet. Use an insulated screwdriver to prevent
troubleshooting tables, or when changing from
accidental shorting to ground.
60-Hz to 400-Hz input power, Proceed as follows:
4-4