Click here to make tpub.com your Home Page

Page Title: PREPARATION FOR USE
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

TM 11-66253017-14
TABLE 2.1.
2.4 PREPARATION FOR USE
Frequency ranges
It is helpful in reading these instructions and in using
the instrument until completely familiar with its operation,
to remember that where a control or switch position is
RANGE switch
Range
Osc. freq.
R.F. input freq.
marked SET, the control should be adjusted to make the
setting
MHz
MHz
meter read on the line marked SET.
1-2
1
5.5 -11
4.0 - 9.5
Before commencing measurements:
2
11 - 22
9.5 - 20.5
3-4
3
22-44
20.5 - 42.5
(1)
Check the mechanical zero of the meter and
4
44.-.88
42.5 - 86.5
adjust, if necessary, by means of the screw.
5-8
5
88-176
86.5 - 174.5
6
176- 352
274.5 - 350.5
(2)
Turn the LEVEL control towards minimum
7
352-704
350.5 - 702.5
(counter-clockwise).
8
701.5-1001.5
700 - 1000
(3)
Connect the signal to be tested to the R. F. IN
EXT
-
F+ 1.5
F
socket at a level which, operating into50 1, will
develop a voltage within the limits given in
(4)
Set the oscillator RANGE switch in accordance
section 1.2 under 'R. F. input' .At the highest
with table 2.1.  For detailed instructions on
carrier frequencies the shortest possible input
oscillator arrangements, see section 2.10.  In
lead should be used or an external 50 f matching
this section, it is assumed that the internal local
attenuator.
oscillator will be used.
CAUTION Inputs greater than 3 V r. m. s. May damage
(5)
Set the Function switch to TUNE OSCILLATOR
the attenuator. If in doubt, use an external attenuator.
AND ADJUST LEVEL. Adjust the OSCILLATOR
TUNE dial to a frequency 1.5 MHz above the
carrier.  Tune for peak meter deflection, and
(4)
Turn the SUPPLY switch clockwise so that the
then adjust LEVEL for a meter deflection in the
red segments show.
black R. F. LEVEL arc, preferably near the top
end for best signal-to-noise ratio.
(6)
Set the Function switch to F.M. SET FREQ.
2.5 MEASURING F.M. DEVIATION
Slightly readjust the OSCILLATOR TUNE control
until the meter deflects to the SET line. Avoid
After carrying out the preparation procedure in
spurious settings - the correct one is the closest
to the original setting obtained in (5).
(1)
Set the A.M. /DEV RANGE switch (below the
(7)
Set the Function switch to DEV + and DEV-.The
meter) to suit the expected deviation of the
demodulated peak frequency deviations above
signal under test.
and below the carrier frequency may be
measured on the meter or viewed on an
(2)
Set the MAX MOD FREQUENCY switch to 200
oscilloscope  connected  to  the  OUTPUT
kHz unless the modulating frequency is less than
terminals.
Read the meter on the scale
15 kHz. If so, switch to 15 kHz in order to obtain
corresponding to the setting of the A. M./ DEV
the best signal-to-noise ratio.
RANGE switch.
(3)
Set the DE-EMPHASIS switch as required. Its
Note
If unequal readings are obtained, distortion in the
setting has no effect on the internal meter
equipment under test is indicated.
reading unless the A.M. /DEV RANGE switch is
in the 1.5 kHz deviation position.  The de-
emphasis is effective on the demodulated signal
(8)
Before carrying out a series of measurements,
at the OUTPUT terminals.
and from time to time during measurements,
2-4

Privacy Statement - Press Release - Copyright Information. - Contact Us

Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business