2-3. SERVICE UPON RECEIPT
Perform the steps outlined in Table 2-1 to determine the condition of the Radiac Set
and any obvious defects.
SERVICE UPON RECEIPT
Action
Item
Location
1) Inspect for rust, fungus, paint damage and
Case
Case
broken handles and clasps.
2) Reject the case if damage prevents it from
functioning properly.
1) Inspect the harness for tears.
Harness
Case
2) Inspect the clips for damage.
3) Reject the harness if damaged.
1) Inspect for dirt, fungus and grease.
Headset
Case
2) Inspect for breaks and cracks in the headset.
Inspect for tears in the cables.
3) Check the headset connector for looseness.
4) Reject the headset if damage will prevent it
from functioning properly.
1) Inspect for grease, dirt and fungus.
Radiacmeter
Probe
2) Inspect for tears in the cable.
3) Reject if the cable is torn.
1) Inspect for grease, dirt and fungus.
Radiacmeter
Case
2) Check for broken glass and cracks on the in-
dicating meter scale window.
3) Check the range switch for looseness or bin-
ding. Check that range switch is not cracked
or missing.
4) Check the captive screws for broken-threads
or gaskets.
5) Check the battery compartment for corrosion
or rust and the rubber gasket for damage.
6) Operational check - refer to TM
11-6665-209-10, Preventive Maintenance
Checks and Services,
7) Reject if indicating meter glass is broken or if
operational check is inconsistent.