TM 11-6625-922-15
(3) Operate SWEEP CALIBRATE switch to
dbm/1% MODULATION meter, for example:
LOW CAL.
(4) Operate the CARRIER FREQUENCY control
TRANSMIT
dbm/%
Output
to the desired low end frequency, monitored on the
LEVEL DBM
MODULATION
level
FREQUENCY display, of the frequency band to be
switch
meter
swept.
(5) Operate the SWEEP CALIBRATE switch to
-10
-3 DB
-13 DBM
HIGH CAL.
0
0 DB
0 DBM
(6) Operate the SWEEP WIDTH control to the
+10
-5 DB
+ 5 DBM
desired high end frequency, monitored on the
FREQUENCY display, of the frequency band to be
m. Operate the RANGE (KHZ) switch to the desired
swept.
frequency range. To operate between 100 Hz and 50
(7) Operate the SWEEP CALIBRATE switch to
kHz, use the .1-50 position. If operating in the 50- to
SWEEP. The output signal will automatically sweep
552-kHz range, use the 50-600 position.
between the frequency limits set up at LOW CAL and
HIGH CAL positions. The rate of the frequency sweep
n. The CARRIER FREQUENCY control is used to
is adjustable by the SWEEP TIME control.
vary the transmit signal within the desired range. This
control can vary the signal out of the ranges, but
p. Connect the input signal to either RECEIVE jack
operation should be maintained within the designated
of the TS-2669/GCM. This is the signal which is
ranges. The reference frequency can also be used by
obtained from the opposite end of the transmission path
operating the SWEEP - CARRIER - REFERENCE
to which the transmit signal is connected. The 75 Ω
switch to REFERENCE and using the REFERENCE
FREQUENCY control as described for the CARRIER
UNBAL RECEIVE jack is used only for 75-ohm
FREQUENCY control. The reference frequency will
impedances in the frequency range from 50 kHz to 552
also be transmitted at the TRANSMIT jacks. It is used
kHz. The BAL RECEIVE jack is used for 130/135-, 150-
as a fixed reference frequency in the band of operation
, 600-, and 900-ohm impedances in the frequency range
which can always be returned to in order to check for
from 0.1 kHz to 110 kHz.
drift in measurements without having to crank the
CARRIER FREQUENCY control all the way back. The
Caution: Do not connect the BAL
CARRIER FREQUENCY control is normally used to
RECEIVE jack directly to a line
make all measurements and the REFERENCE
having a dc potential. Use external
FREQUENCY control is left at a fixed frequency.
coupling
capacitors
for
these
connections.
o. Operate the TS-2669/GCM in the sweep mode
for the signal frequency as follows:
q. Operate the RECEIVE IMPEDANCE Ω switch to
the proper position to match the transmission line being
(1) Operate SWEEP-CARRIER-REFERENCE
used.
switch to SWEEP.
r.
Operate the % MOD-RECEIVE LEVEL-
(2) Operate RANGE (KHZ) switch to desired
TRANSMIT LEVEL switch to RECEIVE LEVEL. Monitor
frequency range of operation.
the level at the RECEIVE jacks on the dbm/%
MODULATION meter.
Figure 3-5. Connection for loop mode.
3-7
|