TM 11-6625-922-15
Control, indicator, or jack
Function
DELAY METER RANGE MS (2-position toggle)
Selects either a coarse or a fine delay reading.
switch
Sw pos
Function
3-10
DELAY meter indicates
coarse delay. For a mod-
ulating frequency of 25
Hz, DELAY meter indi-
cates within a range of
10 milliseconds; 83-1/3
Hz, a range of 3 ms;
250 Hz a range of 1 ms.
15-15
DELAY meter indicates fine
delay. For modulating
frequency of 25 Hz, there
is a range of 0.5 MS
(500 us); 83-1/3 Hz,
0.15 MS (150 us); 250
Hz, 0.05 ms (50 us).
COARSE DELAY MS (4-position rotary) switch
Selects the particular coarse range for DELAY meter
reading.
FINE DELAY MS (20-position rotary) switch
Selects the particular fine range for DELAY meter
reading.
DELAY meter
Indicates envelope delay of an input signal, within a
time range in milliseconds, as determined by settings
of the 3 previously mentioned switches, and the se-
lected modulating frequency. Red numbers are used
with a modulating frequency of 83-1/3 Hz. When
DELAY METER RANGE MS switch is in 3-10 posi-
tion, correct delay is determined by adding algebrai-
cally the number setting of COARSE DELAY MS
switch and DELAY meter indication. When DELAY
METER RANGE MS switch is in .15-.5 position cor-
rect delay is determined by adding algebraically the
number setting of COARSE DELAY MS switch,
the number setting of FINE DELAY MS switch,
and the DELAY meter indication.
DELAY ZERO ADJ control
Adjusts signal delay within instrument, so that relative
delay of a reference signal can be read as zero, or
other convenient number.
REF RETURN-END TO END (2-position toggle)
Selects mode of operation.
switch
RANGE (KHZ) (2-position rotary) switch
Selects frequency range for carrier and reference fre-
quency.
CARRIER FREQUENCY control
Adjust output signal carrier frequency.
REFERENCE FREQUENCY control
Adjust reference carrier frequency.
SWEEP-CARRIER-REFERENCE (3-position rotary)
Sw pos
Function
switch
REFERENCE
Frequency of transmitted
signal is determined by
REFERENCE FREQUENCY
control.
CARRIER
Frequency of transmitted
signal is determined by
CARRIER FREQUENCY
control.
SWEEP
Frequency of transmitted
signal is determined by
CARRIER FREQUENCY
control, plus SWEEP con-
trols. In this position,
SWEEP CALIBRATE
switch is operative.
3-2