cilloscope DELAY TIME MULTIPLIER control
11 controls to position pulse for pulse width meas-
clockwise to position them. The pulses should be ob-
urement.
served in the order given.
f. Measure video pulse width. It should be 0.7
NOTE
+0.0 -0.05 sec.
If necessary, adjust the indicated control
NOTE
to center the pulse.
If necessary, adjust module All pulse
width control A11R6 to obtain the re-
Pulse
Control
Pulse
Control
A1 . . . . . . . . . . . ATR2
D1 . . . . . . . . . . . A7R8
quired performance standard.
C 2 . . . . . . . . . . . . . A7R3
B2 . . . . . . . . . . . . . A7R9
g. Disconnect cable from test set test point
A2. . . . . . . . . . . . . A7R4
D2. . . . . . . . . A7R10
A11TP1 and connect to test point A11TP2. Width
C4 . . . . . . . . . . . . . .A7R5
B4 . . . . . . . . . . . . A7R11
of comparison pulse should be 0.5 0.01 micro-
A4 . . . .. . . . . . . . . . A7R6
F2 . . . . . . . . . . A7R13
second.
B1 . . . . . . . . . . . . . . A7R7
NOTE
i. Set the AN/UPM-98 SIF CODER unit FUNC-
If necessary, adjust comparison Pulse
TION control to ID.
width control A11R15 to obtain the re-
j. Set the test set FUNCTION control to I/P and
the MODE control to position 2.
quired performance standard.
k. Adjust the oscilloscope DELAY TIME MULTI-
7-8. Comparing Pulse Position Test
PLIER control clockwise until the pulses are ob-
Connect the equipment as show in figure 7-1. Set
served. Determine whether the I/P pulses are.
and adjust the controls as described in paragraph
aligned as in g above.
NOTE
a. Disconnect the oscilloscope CHANNEL B cable
If necessary, adjust control A7R14 to cen-
from test point A9TP4 and connect the cable to test
ter the CHANNEL B pulse within the
point A11TP2.
CHANNEL A pulse.
b. Disconnect the CHANNEL A cable from the
AN/UPM-98, and connect the cable to teat point 7-8.1. Comparing Pulse Position Test
Using AN/UMP-137A
A11TP1.
control at B INTENSIFIED by A.
Set and adjust the controls as described in para-
sec and TIME BASE B TIME/CM control to 0.2
a. Connect AN/UPM- 137A oscilloscope
sec.
VERTICAL-CHAN A-VIDEO IN jack to test set test
e. Adjust the DELAY TIME MULTIPLIER con-
point A11TP1 and AN/UPM-137A oscilloscope
trol until the second pulse of the train on CHAN-
VERTICAL-CHAN B-VIDEO IN jack and test set
NEL A is intensified.
test point A11TP2.
f. Set the oscilloscope HORIZONTAL DISPLAY
control at A DEL'D BY B. The C1 pulse of each
as necessary to observe C1 pulse of both displays.
train should be 0bserved.
c. Determine whether pulse displayed on
NOTE
AN/UPM-137A oscilloscope CHAN A is centered
If necessary, carefully adjust the DELAY
with respect to pulse at CHAN A.
TIME MULTIPLIER control to center
NOTE
these pulses on the display.
If necessary, adjust control A7R1 in
g. Determine whether the pulse at CHANNEL B
module A7 to center the pulse as accurate-
is centered with respect to the pulse at CHANNEL
ly as possible.
A.
d. Repeat b and c above for following pulses and ad-
NOTE
just indicated control, if necessary. Pulses should be
If necessary, adjust control A7R1 in
observed and adjusted in order given.
module A7 to center the pulse as accurate-
Pulse
Control
Pulse
Control
ly as possible.
D1 . . . . . . . . . . . .. A7R8
A1 . . . . . . . . . . . . .. A7R2
h. Determine as in g above, whether each of the
B2 . . . . . . . . . . . .. A7R9
C2 . . . . . . . . . . . . .. A7R3
D2 . . . . . . . . . . . .. A7R10
A2 . . . . . . . . . . .. A7R4
pulses listed below is centered by adjusting the os-
7-7