|
|
![]() TM 11-6625-3017-14
SECTION 5
REPAIR
(iii) Systematic calibration, which can be used to locate
5.1 INTRODUCTION
a fault for which methods (i) and (ii) are inadequate
- see Sect. 5.4.
This chapter contains information for the
localization and repair of faults. Performance limits
quoted are for guidance only and should not be taken as
5.2.1
Test equipment required
guaranteed performance specifications unless they are
also quoted in the Data Summary section.
See Table 5.1.
CAUTION
See Maintenance, Sect. 4.1, for
precautions in handling semiconductors and for
5.2.2
Front panel checks
advice on screw fasteners.
In the case of any difficulty, please write to or phone
As a first step in fault finding, the three following
the Marconi Instruments Service Division (see address
tests will help you to verify nearly all the circuits of the
on back cover) or nearest representative, quoting the
modulation meter and to localize the fault to certain
type and serial number on the data plate at the rear of
areas. These tests can be performed without removing
the instrument. If the instrument is being returned for
the cover. Failure of all three tests may indicate a faulty
repair, please indicate clearly the nature of the fault or
power supply, in which case first check the fuses on the
the work you require to be done.
rear panel.
(1)
Turn the Function switch to SET CAL. If the meter
reads SET, the calibrator (A12) and the
5.2 FAULT LOCATION
discriminator, 1st and 2nd 1. f. amplifier, peak
reading voltmeter and associated circuits (A5 to
General procedure
A10) are working. If there is no meter indication,
but the 2 kHz calibrator square wave can be seen
Methodical fault location can be performed at the
with an oscilloscope at the OUTPUT terminals, then
following three levels:
the peak reading voltmeter circuit, (A10), is faulty.
(i)
Front panel functional checks for
general
(2)
Turn the Function switch to TUNE OSCILLATOR
localization of a fault - see Sect. 5.2.2..
AND ADJUST LEVEL. Connect a signal to the R.
F. IN socket and check that, by adjusting the
(ii)
Internal checks for more detailed localization -see
TABLE 5.1
17.
Table 5.1
Test Equipment Required for Fault Location
Item
Type
Minimum specification
Recommended model
a
Signal
Frequency: .1.5 MHz and between
TF 995B2,
4 and 1000 MHz
TF 2006
Output: c. w., up to 100 mV.
b
20 V d. c. at 20 kS/V
Hewlett-Packard
250 V a.c.
HP410A
c
Measuring
Bandwidth:5 MHz
Hewlett-Packard
oscilloscope Sensitivity: 20 mV/cm
HP 1700B
5-1
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |