|
|
TM 11-6625-3017-14
SECTION 4
MAINTENANCE
units above and below the chassis. Components may be
4.1 INTRODUCTION
identified by markings on the boards and chassis. For
replacement of sub-assemblies and components, see
This section contains information for keeping the
equipment in good working order and for checking its
overall performance.
4.3 PERFORMANCE CHECKS
The tests in this section may be used as a routine
CAUTION This instrument uses semiconductor
maintenance procedure to verify the main performance
devices which, although having inherent long term
parameters of the instrument.
All tests can be
reliability and mechanical ruggedness, are susceptible to
completed without removing the case, except where
damage by overloading, reversed polarity and excessive
some internal readjustment is indicated. Tests may be
heat or radiation. Avoid hazards such as reversal of
done at any convenient frequency within the carrier
batteries, prolonged soldering, strong r. f. fields or other
range of the modulation meter provided that this
forms of radiation, use of insulation testers or
frequency is also within the range of the test equipment
accidentally applied short circuits. Even the leakage
items a, b, c, d and g.
current from an unearthed soldering iron could cause
trouble. Before shorting or breaking any circuit, refer to
the circuit diagrams to establish the effect on bias
4.3.1
Test equipment required
arrangements of the transistors.
See Table 4. 1.
Screw fasteners
4.3.2
R.F. input and i.f. output
Screw threads used on this instrument are of the
following sizes - 8BA, 6BA, 4BA, 2BA and 1/4 BSF.
Test equipment:- a, d, e and 1.
Cruciform headed screws are of the Phillips
Pozidriv pattern; to avoid damaging them a Pozidriv
This procedure verifies the sensitivity of the
screwdriver should be used.
modulation meter and the adjustment of the local
4.2 ACCESS AND LAYOUT
R.F. input resistance
The main case assembly is held in position by the
rear panel. This panel may be removed after unscrewing
(1)
Turn the LEVEL control mid-way. Connect the
the four coin-slotted screws across the corners; the
multimeter, set for resistance measurement,
mains lead is held to the rear panel by a clip and will
between the centre pin of the R. F. IN socket and
earth and check that the reading is 50 Ω 10 Ω
come away with it if the mains plug is withdrawn. The
instrument can now be slid forward, out of the case;
alternatively, it can be placed face-downwards on a soft
surface and the case lifted off.
R.F. sensitivity
Rack-mounted models have top and bottom dust
(2)
Adjust the signal generator to give a c. w. output of
covers, each attached by four screws.
about 50 mV and connect it to the R. F. INPUT
socket of the modulation meter and to the
voltmeter, keeping the leads as short as possible.
4-1
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |