1-11. GENERAL
Radiac Set AN/PDR-27(#) is designed to detect beta radiation and measure and detect
gamma nuclear radiation. The AN/PDR-27(#) is used to monitor low level radiation
contamination on personnel and equipment. The highest reading (500mR/hr.) is ex-
tremely low compared with other readings likely to be encountered on the nuclear
battlefield.
The AN/PDR-27(#) functions as shown on the accompanying block diagram.
RANGE SWITCH. Activates batteries and controls sensitivity of meter.
A.
BA-30 BATTERIES. Supply operating power to the radiacmeter.
B.
POWER SUPPLY CIRCUITS. Supply required voltages to the tubes in the probe and
C.
to the amplifier and pulse shaping circuit.
REGULATOR CIRCUITS. Keep power supply circuit voltages constant so that
D.
radiacmeter operates properly.
PROBE. Contains two Geiger-Mueller tubes which detect radiation. One tube is
E.
associated with range switch position 500 to 50, the other with position 5 or 0.5.
AMPLIFIER AND PULSE SHAPER CIRCUITS. Takes pulses from the probe and com
F.
verts them to constant amplitude pulses which are applied to the meter circuit.
METER. Indicates the amount of radiation detected by the probe on a scale.
G.
HEADSET. Provides an audible indication of radiation detected by the probe.
H.
|