|
|
TM 11-6625-667-45/NAVAIR 16-30APM123-2/TO 33A1-3-367-22
CHAPTER 4
REPAIRS AND ALIGNMENT
4-1. General Parts Replacement
soldering gun; damaging voltages can be induced
Techniques
in components.
Most of the parts of Test Set, Transponder
quickly wherever wiring permits, use a heat sink
AN/APM-123V are readily accessible since they
(such as a pair of long nosed pliers between the
are located on pullout circuit boards. Access to
s o l d e r e d joint and the transistor. Use ap-
parts in the transmitter and 28-volt power
p r o x i m a t e l y the same length and dress of
supply, h o w e v e r , requires removal of its
transistor leads as used originally.
respective chassis from the test set for removal
and replacement. All front panel switches require
4-2. Adjustment
removal of one or more subassemblies for access
NOTE
to repair or replace them. After repairing the
4-6 when
See para
using
equipment, the adjustments in paragraph 4-2
AN/TPM-25A.
should be performed. The precautions in a
through e below apply specifically to this test set.
After a circuit in a fictional section is repaired,
a. Do not disturb the adjustment of preselector
the functional section should be adjusted to
A15Z1 (fig. 3-6). It was adjusted at the factory.
insure proper equipment operation. Two ad
This adjustment is extremely critital, and special
justment procedures are provided for the tran-
test equipment is required to test it or to perform
smitter, receiver and decoder functional sections
an adjustment. Adjustment of this preselector
to use either conventional test equipment and the
will void the entire calibration of the unit. If this
Radar Test Set AN/UPM-98 or the Radar Test
part appears faulty, it should be replaced with a
Set AN/UPM-137A. The functional sections are
new one.
listed below together with the reference to the
b. Do not disturb the settings of controls A5R1
paragraph containing the adjustment procedures.
through A5R6 and A7R1 through A7R14 (fig.
factory a n d are extremely critical. Any
movement of one or more controls will void the
entire calibration of the unit.
4-5.1)
CAUTION
4-3. Power Supply Adjustments
If control A16A2R1 of the transmitter
a. Test Equipment Required. Multimeter
voltage regulator is adjusted to a
ME-26B/U is required.
v o l t a g e greater than specified it will
degrade and shorten the transmitter
electron tube performance and life.
below. Refer to figure 3-6 for test point location
c. Do not adjust control A16A2R1 (fig, 8-25) of
and figure 3-7 for adjustment control location.
the transmitter voltage regulator without first
c. Procedure
r e f e r r i n g to the instructions o u t l i n e d in
(1) Set the ME-26B/U SELECTOR control
to(+) and RANGE control to 300V.
calibration of this equipment.
(2) Connect the ME-26B/U common lead to
test point A13TP2 (ground and the dc probe to
watt maximum capacity. This test set is tran-
test point A14TP3.
sistorized. If the soldering iron must be used with
(3) Apply power to the test set.
alternating current (ac, use an isolating trans-
(4) Adjust A14R8 until the meter indicates
former between it and the line. Do not use a
+ 150 2 volts.
Change 5 4-1
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |