Control settings
Paragraph
Normal Indication
Equipment under test
Test equipment
Procedure
reference
Step
Same as step 1.
TS 779A/U
CAUTION
4
BOLO BIAS CURRENT
TS 779A/U BOLO BIAS CURRENT
OFF
control must be set to OFF before the
BOLO RES 200
MX-2144A/U is connected to the
BOLO TEMP COEF:
TS-779A/U to avoid equipment
damage
NEG
a. Connect the MX-2144A/U to the TS 779A/U a. None.
input.
b. None.
b. Adjust the TS 779A/U as follows:
(1) RANGE switch at 5 DBM.
(2) BOLO BIAS CURRENT switch at 10-16
.
(3) LINE POWER switch at ON and permit a
15-minute warmup period.
(4) Adjust ZERO SET COARSE and FINE
controls to zero meter.
A13TP2.
d. Connect the MX-2144A/U to the test set d. 9 dbm 1.
PROBE jack. Observe the power in
dication.
e. Disconnect the ground from A4TP5 and con- e. None.
nect A4TP4 to ground.
f. Observe the power meter indication.
f. -21 dbm 1.
AN/UPM-137 Same as step . . . . . . . . .
5
VERTICAL CHAIN A-
VOLTS DIV: .5
Permit a few minutes warmup time.
VERTICAL CHAN A-
connect oscilloscope to test set A4TP6.
ACIDC:AC
VERTICAL CHAN A-
VERTICAL, INTENSITY, SWEEP
DELAY, FOCUS, and ASTIGMATISM
OUT
controls until a clear waveform is ob-
SYNC:EXT AC
served.
c. Set the transponder test MODE switch to 1, c. Pulse pair (P1-P3) and P2
2, 3A, and C and observe whether a pulse
pulses present for model
pair (P1 P3) and low level P2 pulse are
1, 2, 3,/A. TEST and C.
present for each mode.
|