Page Title: Table 6-1. Troubleshooting-Continued
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

TM 11-6625 -537-14-1
Table 6-1. TroubleshootingContinued
Corrective action
Probable cause
Malfunction
a. Faulty resistor in calibration
a. Check and replace faulty resistor.
32. REF CAL ADJ R318 cannot be
sampling string R316 through
adjusted to null VOLTS meter.
R319.
b. Faulty standard cell R1
b. Check standard cell for between
1.0190 and 1.0196 volts. Replace
if voltage is out of tolerance.
CAUTION
Use only a null type
meter to avoid damaging
the standard cell. The
standard cell is isolated
from all internal circuits
by  the  OPERATE-
CALIBRATE  switch,
The plug to the standard
cell should be removed
before performing any
voltage or resistance
measurements on the
OPERATE-CALI-
BRATE switch to pre.
vent accidental shorting
and damage to the cell,
NOTE
Normal- service life of the
standard cell is 8 to 15
years. End of service life
is usually marked by an
increase in temperature
hysteresis effect. That is,
reading errors in excess
of 0.025 percent will
result when the same
voltage is read when the
differential voltmeter is
cold  and at  normal
operating temperature
6-6