TM 11-6625 -537-14-1
string. Between the two wipers of S6 then, there
VDC ADJ R329 adjust the output level for the 50,
is a total resistance of 40K (80K paralleled by
5, and 0.5-volt ranges, respectively, In the ac
80 K). If 500 volts is applied to the precision
mode of operation, levels G and H of AC-de
voltage divider, a total of 100 volts will then
polarity switch S5 bypass RANGE switch S2 and
appear across the wipers of S6, 10 volts across the
fix the output of the reference voltage range
wipers of S7, 1.0 volt across the wipers of S8, O..
divider at 5 volts.
volt across the wipers of S9, and 0.01 volt across
each resistor selected by S1O. The wirewound
precision voltage divider divides the output of the
resistors comprising the precision voltage divider
reference voltage range divider (500, 50, 5, or 0.5
are matched to provide an overall divider ac-
volts) into 500 equal increments. The precision
curacy of 0.01 percent. A filter consisting of R331
voltage divider consists of five Kelvin -Varley
and C301 or C302 is connected across the output
decade resistor strings, R401 through R449, and
of the precision voltage divider by level D of
switches S6 through S1O. The position of
RANG E switch S2 to eliminate any electrical
switches S6 through S 10 is indicated by voltage
noise introduced by the wirewound resistors in
readout dials A through E, respectively, on the
front panel. The strings connected to S7, S8, S9,
t h e precision voltage d i v i d e r and reference
and S 10 parallel two resistors of the preceding
voltage range divider,