TM 11-6625-537-14-1
c. Operating Temperature Range.
1-8. Description
(1) Within dc specifications from 40 F.
The Electric Voltmeter ME-202(*)/U is a single
to 105 F. De-rate accuracy at 0.001 percent per
chassis unit enclosed in a metal case. The meter,
F. outside these limits to 35 F. and 110 F.
o p e r a t i n g controls and input connectors are
(2) Within ac specifications from 55 F. to
mounted on the front panel. The adjustment of
95 F. De-rate accuracy at 0.005 percent per 0 F.
the front panel controls adjusts a precision dc
outside these limits to 35 F. and 110 F.
internal voltage to equal the external voltage
being measured; equality is indicated by a null on
CAUTION
the panel meter; ac voltages are first converted to
The standard cell may freeze if subjected
a dc voltage for measurement. The power fuse,
t o below freezing temperature. The
recorder output connectors, recorder gain and
e l e c t r o l y t e will freeze at 1 F. and
adjustment,  a n d the power supply cord are
operation below 32 F. is definitely not
located on the back panel.
recommended.
d. Calibration. The 500-volt dc reference supply
1-9. Tabulated Data
is calibrated against a built-in standard cell.
a. Technical Characteristics. The technical data
e. Input Power. 115/230 vac, 50 to 440 Hz, 75
f o r the Electronic Voltmeter ME-202(*)/U in-
watts.
eluding ranges, accuracy and tolerances are listed
f. Number of Tubes. 8.
in tables 1-1 through 1-7.
g. Weight. 28 l b s .
b. Stability of Reference Supply.
1-10. Items Comprising an Operable Equipment
(1) 0.0025 percent for 10 percent change in
line voltage.
Electronic Voltmeter ME-202(*)/U comprises an
(2) 0.005 percent per hour after 30-minute
operable equipment.
w a r m u p (adjustable front panel CALIBRATE
control).
Table 1-1. Conventional Vacuum Tube Voltmeter Specifications*
Scale
Input impedance
Range
500-0-500
0 to 500 Vdc
50 megohms
50-0-50
0 to 50 Vdc
50 megohms
10 megohms
10-0-10
0 to 10 Vdc
5-0-5
0 to 5 Vdc
50 megohms
1-0-1
10 megohms
0 t0 1Vdc
0.5-0-0.5
50 megohms
0 to 0.5 Vdc
0.1-0-0.1
10 megohms
0 to 0.1 Vdc
0.01-0-0.01
1 megohm
0 to 0.01 Vdc
0-500
1 megohm 35 f
0 to 500 Vac
0-100
1 megohm 35 f
0 to 100 Vac
0-50
1 megohm 35 f
0 to 50 Vac
0-10
1 megohm 35 f
0 to 10 Vac
0-5
1 megohm 50 f
0 to 5 Vac
0-1
1 megohm 35 f
0 to 1 Vac
1 megohm 50. f
0-0.5
0 to 0.5 Vac
0-0.1
1 megohm 35 f
0 to 0.1 Vac
0-0.01
1 megohm 50 f
0 to 0.01 Vac
0-0.001
1 megohm 50 f
0 to 0.001 Vac
* Accuracy: 3 percent on all ranges
1-2

Privacy Statement - Press Release - Copyright Information. - Contact Us

Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business