TM 3-4240-201-13
Section III. OPERATION OF THE FILTER UNIT
disconnects the 24 volt dc power supply. The centrifugal
2-7.
Power Supply
fan operates on 24 volts dc or 30 volts ac.  The
centrifugal fan blows the air to the six headpieces
a. Transformer. The transformer, a part of the filter
through the connected hoses.
unit, receives the external power supply. The 115 volt 60
d. 115 Volt 60 Hz ac Power Cable. The 115 volt 60
Hz ac power supply and the 24 volt dc power supply are
Hz ac power cable is 6 feet long. It is a two wire rubber
not part of this filter unit.
covered cable that connects into ordinary housepower
b. General. Make sure that the toggle switch on
outlets by plug connector.
the transformer is in OFF position. Make sure that the
e. 24 Volt dc Power Cable. The 24 volt dc power
power cable that extends from the transformer group is
cable is 3 feet long. It is a two wire rubber covered cable
connected to the air purifier.  Then connect the two
that has two clips covered with one black and one red
power cables that extend from the transformer as
plastic cable nipples. Squeezing the cable nipples opens
follows:
the spring-loaded clips. Clamp the black clip over the
(1) to the 115 volt 60 Hz ac power source.
negative terminal and the red clip over the positive
(2) to the 24 volt dc power source. The black
terminal.
cable nipple and clip connect to the negative terminal of
f. Relay. The relay is a double pole double throw
the battery and the red cable nipple and clip connect to
relay that operates when 115 volt 60 cycle ac power is
the positive terminal of the battery.
applied to the armature.
The relay is inside the
transformer case and it operates automatically when the
NOTE
toggle switch is on with 115 volts ac power applied.
Two 12 volt batteries must be connected
g. Air Pitrifier Cable. The air purifier cable is 3 feet
in series as shown in figure 1-1 in order
90
long. It is a two wire rubber covered cable with a
to get the required 24 volt dc power
electrical connector. The air purifier cable connects the
supply as shown in the wiring diagram
transformer to the air purifier.
(fig. 18).
2-8.
Protective Masks
c. Electrical Operation. When the toggle switch is
turned on and the 115 volt power cable is connected to a
A protective mask will be required for each operator of
power source, current flows through the primary winding
the filter unit. The operator may be referred to as nurse,
of the step down power transformer and through the
orderly, attendant, doctor, medic, aid man, or operator.
armature of the relay, actuating the relay.
The
The operator is responsible for first donning a protective
secondary winding of the step down power transformer is
mask; and then putting headpieces on six patients. A
connected through the toggle switch to the centrifugal
headpiece is not intended to be worn by the operator
fan in the air purifier. If the 115 volt ac power source
(attendant) of the filter unit. The protective mask
ceases for any reason, current ceases to flow through
required for the operator is not part of this filter unit. The
the armature of the relay, and the relay contacts open.
protective mask required for operator protection is the
The 24 volt dc power source is connected to the
authorized field protective mask.
centrifugal fan through a second set of contacts that are
closed in the relay and through the toggle switch. If at
any time during the operation 115 volt ac power
becomes available, the relay connects the  secondary
windings of the transformer to the centrifugal fan and
Section IV. OPERATION UNDER UNUSAL CONDITIONS
2-9.
General
2-10.
Details
a. Keep the filter unit free of sand, moisture, mud,
Operation under unusual conditions includes cold
weather, wet weather, mud, seaside, high humidity, and
or any other foreign matter.
high temperatures. Protect the filter unit from exposure
to any of these conditions as much as possible.
2-3