SB 740-91-01
StepSetting and InstructionsNormal Indication
2(7) Adjust AN/URM-103 LO RF pV control
(cont.)for minimum output.
(8) Check to see that AN/URM-103 Deviation
Meter still indicated 3 kHz. Adjust
deviation adjust control, if necessary.
(9) Slowly advance AN/URM-103 control untilLO RF lV control set-
Call indicator lights. Note setting ofting should be 0.5 pV
LO RF lV control.or less.
(10) Slowly reduce inputs to 0.4 V. Tone(Call indicator remains
squelch should remain broken.lighted.)
(11) Repeat for frequency of 64.05 MHz by
adjusting the AN/URM-103 controls as
follows:
(a) Band Switch. D
(b) RF Tuning. 64.05 MHz
(12) Check to see than AN/URM-103 Deviation
Meter still indicates 3 kHz. Adjust
deviation adjust control, if necessary.
(13) Slowly advance AN/URM-103 control untilLO RF lV control set-
Call indicator lights. Note setting ofting should be 0.5 liV
LO RF PV control.or less.
3Audio output and distortion test
(a) Connect the equipment in accordance
with figure B-8.
(b) Energize all equipment and allow for
a 5 minute warm-up before proceeding.
(c) On AN/USM-207, set controls as follows:
(1) Display Control. Set for desired
display time.
(2) Sensitivity. 100 V
(3) Gate Time. (sec-L):106
(4) Function. Freq
(d) On AN/URM-103, set controls as follows:
(1) Band Switch. D
(2) RF Tuning. 60 MHz
(3) Deviation Range. 10 kHz
(4) Function. 1000 Hz
(5) Deviation Level. Set for 8 kHz
on Deviation Meter.
(6) LO RF Set to Line. Right-hand
meter to red line.
(7) RF Output. LO
B-48
|