Click here to make tpub.com your Home Page

Page Title: Read Delay, Error Detector, and Function (Controls) Test
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

TM 11-6625-667-45/NAVAlR 16-30APM123-2/TO 33A1-3-367-22
DISPLAY control at "B" INTENSIFIED BY
k. Set the test set FUNCTION control to I/P
and the MODE control to position 2.
"A".
e. Set the TIME BASE A TIME/CM control to
l. Adjust the oscilloscope DELAY TIME
0.1 s and TIME BASE B TIME/CM control to
MULTIPLIER control clockwise until the pulses
0.2 s.
are observed. Determine whether the I/P pulses
f. Adjust the DELAY TIME MULTIPLIER
are aligned as in h above.
control until the second pulse of the train on
NOTE
CHANNEL A is intensified.
If necessary, adjust control A7R14 to
g. Set the oscilloscope HORIZONTAL
center the CHANNEL B pulse within
DISPLAY control at "A" DEL'D by "B". The
the CHANNEL A pulse.
Cl pulse of each train should be observed.
7.1-10. Read Delay, Error Detector, and
NOTE
Function (Controls) Test
If necessary, carefully adjust the
DELAY TIME MULTIPLIER control
a. Perform the initial setup procedures given in
to center these pulses on the display.
h. Determine whether the pulse at CHANNEL
b. Connect the oscilloscope CHANNEL A
B is centered with respect to the pulse at
cable to test point A10TP6 (module A10).
c. Measure to make sure that the width of the
CHANNEL A.
pulse at CHANNEL A is 150 7.5 s.
NOTE
d. Disconnect the oscilloscope CHANNEL A
If necessary, adjust control A7R1 in
cable from test point A10TP6, and connect the
module A7 to center the pulse as ac-
cable to test point A11TP3.
curately as possible.
e. Set the oscilloscope  HORIZONTAL
i. Determine, as in h above, whether each of the
DISPLAY control to "B" INTENSIFIED BY
pulses in the chart below is centered by adjusting
"A", TIME BASE A TIME/CM control to 0.1
the oscilloscope DELAY TIME MULTIPLIER
s and TIME BASE B TIME/CM control to 0.2
control clockwise to position them. The pulses
s.
should be observed in the order given.
f. Adjust the DELAY TIME MULTIPLIER
NOTE
control until the pulse at the oscilloscope
If necessary, adjust the indicated
CHANNEL A is intensified.
control to center the pulse.
g. Set the HORIZONTAL DISPLAY control
to "A" DEL'D by "B".
Pulse
Control
h. Measure to make sure that the width of the
Al . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R2
pulse at oscilloscope CHANNEL A is 0.35 0.01
C2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R3
s.
A2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R4
C4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R5
.
NOTE
A4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R6
If necessary, adjust control A11R30 to
B1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R7
obtain the required performance
D1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R8
B2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R9
standard.
D2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R10
i. Set the test set and AN/TPM-25A controls
B4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R11
for each mode as indicated in the chart below. The
D4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. A7R12
F2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A7R13
AN/TPM-123 ACCEPT indicator should light
for each of the modes.
j. Set  the  AN/TPM-25 REPLIES
MODULATION SEL switch to I P M2/3.
Change 5
7.1-5

Privacy Statement - Press Release - Copyright Information. - Contact Us

Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business