TM 11-6625-542-14-1
(2) Remove diode which couples defective
(4)  If the input signal to the multivibrator
stage to following stages.  If the defective stage then
registers on the oscilloscope but no signal appears at the
functions normally, the trouble is probably due to loading
output, measure the voltage at multivibrator plate which
by the stages that follow.
connects directly to the input diode plate. The voltage
reading should be +225 volts. The plate voltage of the
NOTE
other half of the tube should read + 100 volts. The grid
Before replacing any components, isolate the
voltage on the first half of the tube should read
trouble by checking the multivibrator with an
approximately --6 to --8 volts.
oscilloscope and voltmeter.
d.
Multiplier Checks. The multiplier circuits may
(3)
To check multivibrator, check the
become faulty due to tube failure. These troubles may
incoming signal at the grid on one-half of the tube and
be corrected by replacing the faulty tube. Readjustments
the output signal. An oscilloscope is used for this test. If
are sometimes necessary due to aging of components
desired signals are present and pin voltages are correct,
and to variations in characteristics of replacement parts.
check for signals at the output cathode follower stage. If
Refer to section III for adjustments.
no signal appears on the cathode, check through switch
S710 to the MARKER OUT connector.
Table 6-2. Troubleshooting
Malfunction
Probable cause
Corrective action
1.
Improper or loss of time-marks when 1
Defective tube V262, V265, V273
Check and replace defective tube.
MICROSECONDS button depressed.
2.  Improper or loss of time-marks when 5
Defective tube V262, V265, V273, V502,
Check and replace defective tube.
MICROSECONDS button is depressed.
V503, or V533.
3.  Improper or loss of time-marks when 10
Defective tube V502, V503, V533, V522,
Check and replace defective tube.
MICROSECONDS button is depressed.
V543, or V535
4.  Improper or loss of time-marks when 50
Defective tube V522, V543, V535, V552,
Check and replace defective tube.
MICROSECONDS button is depressed.
V553, or V583.
5. Improper or loss of time-marks when
Defective tube V552, V553, V533, V572,
Check and replace defective tube.
100 MICROSECONDS button is
V593, or V585.
depressed.
6.  Improper or loss of time-marks when
Defective tube V572, V593, V585, V402,
Check and replace defective tube.
500 MICROSECONDS button is
V403, or V433.
depressed
7.  Improper or loss of time-marks when 1
Defective tube V402, V403, V433, V422,
Check and replace defective tube.
MILLISECONDS button is depressed.
V443, or V435.
8.  Improper or loss of time-marks when 5
Defective tube V422, V443, 435, V452,
Check and replace defective tube.
MILLISECONDS button is depressed.
V453, or V483.
9.  Improper or loss of time-marks when 10
Defective tube V452, V453, V483, V472,
Check and replace defective tube.
MILLISECONDS button is depressed.
V493, or V485.
10. Improper or loss of time-marks when 50
Defective tube V472, V493, V485, V302,
Check and replace defective tube.
MILLISECONDS button is depressed
V303, or V333.
11. Improper or loss of time-marks when
Defective tube V302, V303, V333, V322,
Check and replace defective tube.
100MILLISECONDSbutton is
V343, or V335.
depressed.
12. Improper or loss of time-marks when
Defective tube V322, V343, V335, V352,
Check and replace defective tube.
500MILLISECONDSbutton is
V353, or V383.
depressed.
13. Improper or loss of time-marks when 1
Defective tube V352, V353, V383, V393,
Check and replace defective tube.
SECONDS button is depressed.
V385, or V372.
14. Improper or loss of time-marks when 5
Defective tube V393, V385, V372.
SECONDS button is depressed.
15. Improper or loss of trigger pulses when
Defective tube V262, V273, V265, V502,
Check and replace defective tube.
100KC-10SCS button is depressed.
V503, or V533.
16. Improper or loss of trigger pulses when
Defective tube V522, V543, V535, V552,
Check and replace defective tube.
10 KC-100 S button is depressed.
V553, or V583.
17. Improper or loss of trigger pulses when
Defective tube V572, V593, V585, V402,
Check and replace defective tube.
1KC-lMS button is depressed.
V403, or V433.
18. Improper or loss of trigger pulses when
Defective tube V422, V443, V435, V452,
Check and replace defective tube.
100 - lOMS button is depressed.
V453, or V483.
19. Improper or loss of trigger pulses when
Defective tube V472, V493, V485, V302,
Check and replace defective tube.
10S-lOOMS button is depressed.
V303, or V333.
20. Improper or loss of trigger pulses when
Defective tube V322, V343, V335, V352,
Check and replace defective tube.
1S-lSEC button is depressed.
V353, or V383.
6-7